Density
- Standard >0.02 g/cm³ (>0.0007 lbs/in³)
- adjustable >0.008 g/cm³ (>0.0003 lbs/in³)
Electromechanical data
Cable entry/plug
3)
- Single chamber housing l 1x cable entry M20x1.5 (cable-ø 5 … 9 mm),
1x blind plug M20x1.5
or:
l 1x closing cap ½ NPT, 1x blind plug ½ NPT
or:
l 1x plug (depending on the version), 1x blind
plug M20x1.5
Spring-loaded terminals for wire cross section up to 1.5 mm² (0.0023 in²)
Adjustment elements
Mode switch
- min. min. detection or dry run protection
- max. max. detection or overfill protection
Voltage supply
Voltage supply 10 … 36 VDC(via the signal conditioning
instrument)
Electrical protective measures
Protection IP 66/IP 67
Overvoltage category III
Protection class II
Approvals
4)
ATEX II 1G, 1/2G, 2GEExiaIIC T6
5)
ATEX II 1G, 1/2G, 2GEExiaIIC T6 + ATEX II 1/2 DIP66 T
ATEX II 1/2G, 2GEExdIIC T6
3)
Depending on the version M12x1, acc. to DIN 43650, Harting, Amphenol-
Tuchel, 7/8" FF
4)
Deviating data with Ex applications: see separate safety instructions
5)
only in conjunction with electronics "Z"
Supplement
38 OPTISWITCH 310 0 C - with two-wire output
29955-EN-060112