32 1260 Nanoflow Pump User Manual
1 Introduction to the 1260 Infinity Nanoflow Pump
Setting the 8-bit Configuration Switch (On-Board LAN)
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
CAUTION
Loss of data
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are diagnosis and repair log books which will not be erased.
Save your methods and data before executing a forced cold start.
Tabl e 12 Forced Cold Start Settings (without on-board LAN)
Mode Select SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
LAN TEST/BOOT11000001
No LAN TEST/BOOT11001001