72 1260 Nanoflow Pump User Manual
5 Optimizing Performance
Hints for the Micro Vacuum Degasser
Hints for the Micro Vacuum Degasser
If you are using the vacuum degasser for the first time, if the vacuum degasser
was switched off for any length of time (for example, overnight), or if the
vacuum degasser lines are empty, you should prime the vacuum degasser
before running an analysis.
The vacuum degasser can be primed by pumping solvent with the 1260
Nanoflow Pump at high flow rate (2.5 mL/min). Priming the degasser is
recommended, when:
• vacuum degasser is used for the first time, or vacuum chambers are empty,
• changing to solvent that are immiscible with the solvent currently in the
vacuum chambers,
• the pump was turned OFF for a length of time (for example over night) and
volatile solvent mixtures are used.
For more information see Agilent 1260 Infinity (G1379B) Micro Degasser User
Manual (p/n G1379-90013).