118 Bio-inert Quaternary Pump User Manual
9 Maintenance
Exchanging the Outlet Valve
1
Using the 1/4 inch wrench disconnect the valve capillary
from the outlet valve. Using the 14 mm wrench loosen the
valve and remove it from the pump body.
2 Do not disassemble the outlet valve, as this can damage
the valve.
NOTE
The bio-inert outlet valve can be identified by a ring,
which is normally covered by the plastic jacket.
3 Reinstall the outlet valve and tighten the valve. Reconnect
the valve capillary.