Directed Electronics G4203P Manual de usuario Pagina 177

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 194
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 176
Bio-inert Quaternary Pump User Manual 177
Hardware Information
12
Setting the 8-bit Configuration Switch (without On-board) LAN
Communication Settings for RS-232C
The communication protocol used in the column compartment supports only
hardware handshake (CTS/RTR).
Switches 1 in down and 2 in up position define that the RS-232C parameters
will be changed. Once the change has been completed, the column instrument
must be powered up again in order to store the values in the non-volatile
memory.
Use the following tables for selecting the setting which you want to use for
RS-232C communication. The number 0 means that the switch is down and 1
means that the switch is up.
The following tables represent the configuration switch settings for the modules without
on-board LAN only.
Tab le 12 8-bit Configuration Switch (without on-board LAN)
Mode Select12345678
RS-232C 0 1 Baudrate Data
Bits
Parity
Reserved 1 0 Reserved
TEST/BOOT 1 1 RSVD SYS RSVD RSVD FC
The LAN settings are done on the LAN Interface Card G1369A/B. Refer to the
documentation provided with the card.
Tab le 13 Communication Settings for RS-232C Communication (without on-board LAN)
Mode
Select
12345 6 78
RS-232C 0 1 Baudrate Data Bits Parity
Vista de pagina 176
1 2 ... 172 173 174 175 176 177 178 179 180 181 182 ... 193 194

Comentarios a estos manuales

Sin comentarios