Practical Example: Dry Weight Determination with Tare Weighing
Menu code settings used in the example:
Function Code
For MC 210 S,
For
MC 210 P,
MC 5 (-0CE),
MC 410 S (-0CE)
SC 2 (-0CE)
and MC 21 S
Differential
weighing and
backweighing 2 1 6 2 1 6
Combined
weighing mode 3 7 3* 3 6 3*
Automatic
data output 7 1 2 7 1 2
With data
ID code 7 2 2 7 2 2
2 decimal
places for the
results in percent 3 6 3* 3 5 3
With tare
weighing 3 8 2 3 7 2*
1st calculated
value: residue
in percent 3 9 2 3 8 2
2nd calculated
value: residue
in wt. unit 310 1 3 9 1
Record/printout
with net initial
sample weight 7 5 3* 7 5 3*
Record/printout
with residue
in wt. unit
and percent 7 6 4 7 6 4
Application: Determination of the dry weight (solids) in percent of samples over
a relatively long period, e.g., using a drying oven. The containers for
the samples are consecutively numbered and placed in the correct
sequence. In this particular application, let’s suppose that your in-
house standard operating procedure does not allow backweighing
to be done immediately after weighing-in a sample.
* = factory setting
As alternatives, you can select
these codes:
Different number
of decimals 3 6 x 3 5 x
Without tare
weighing 3 8 1 3 7 1*
Different calcu-
lated value 3 9 x 3 8 x
Different calcu-
lated value 3 10 x 3 9 x
Other record
printout/output 7 5 x 7 5 x
Other record
printout/output 7 6 x 7 6 x
3–25