Activation of the Data Output Port Lines
You can have different data output port lines – provided by the balance interface –
activated according to your requirements by setting the appropriate menu
code listed below.
For filling, the port lines can be activated regardless of stability. However, for
standard checkweighing applications, the port lines should be activated only
at stability. This increases the reading speed for an externally connected
checkweighing display (YRD 10 Z).
The checkweighing range is within approx. 30% and 170% of the target weight.
Activate the port lines as follows: Code
Only within the checkweighing range 4 3 1*
Always 4 3 2*
Only at stability within checkweighing range 4 3 3
Only at stability 4 3 4
Use in Combination with Other Programs
The over/under checkweighing programs can be used in combination with other
application programs. This means that you will have all functions of the combined
program available in addition.
Program Code
For MC 210 S, For MC 5(-0CE),
MC 210 P, SC 2 (-0CE)
MC 410 S (-0CE) and MC 21S
Counting 2 1 4 2 1 4
Weighing in percent 2 1 5 2 1 5
Differential weighing and backweighing 2 1 6 2 1 6
Density 2 1 7
not applicable
Diameter determination not applicable 2 1 7
Air buoyancy correction program 2 1 8 2 1 8
* = factory setting; depends on the balance model in some cases
3–54