Practical Example: Counting Small Parts
Menu code settings used in the example:
Function Code
Counting 2 1 4
Ref. quantity and ref. weight 7 1 2
With data ID code 7 2 2
Application: Counting of bulk quantities of items with the same weight and with a
light average piece weight
Step/Key ID/Key Readout Data Output
Place container on pan +
122.650 mg
c, t
0.000 mg
Add 20 parts;
enter reference sample
quantity: 20 +
20 pcs
Store:
STO +
20 pcs
nRef + 20 pcs
REF - F wRef + 5.6546 mg
To update ref. sample, add
more parts (approx. 2x nRef): +
38 pcs
OPT - F +
5.6591 mg
wRef + 5.6591 mg
Fill container with desired
quantity of parts +
500 pcs
p +
500 pcs
Qnt + 500 pcs
Toggle to weight readout: w +
2829.550 mg
p +
2829.550 mg
N + 2829.550 mg
The data ID codes mean:
nRef + 20 pcs reference sample quantity (piece count)
wRef + 5.6546 mg reference weight (average piece weight)
Qnt + 500 pcs calculated piece count
N + 2829.550 mg net weight
As an alternative, you can select this
code:
Reference weight only 7 1 3
3–57