Multiple Calibration Mode
The calibration value is calculated from the average of the individual calibration
procedures. You can use the “multiple calibration mode” for both internal
and external calibration. The number of calibration procedures is indicated in the
3rd place of the application display field (e.g., “C.I.3”).
Important Note:
If “Err 04” is briefly displayed, the calibration values deviate too much
from one another. This means the calibration value measured is not stored,
and the calibration procedure will be repeated.
Multiple calibration mode Code
Off 1 14 1*
On 1 14 2
Self-Calibration “isoCAL”and Linearization
Code
Off 1 15 1
Calibration status displayed only
1
) 1 15 2
Self-calibration “isoCAL” on 1 15 3*
Self-calibration “isoCAL” and linearization on 1 15 4
1
) = The symbol “
W
” flashes in the display until you press the appropriate key
to activate one of the calibration functions
CAL Using f
You can activate the “internal calibration” function anytime at the touch
of the f key (factory setting).
Function of the f key Code
Acces denied 2 2 1
Internal calibration “CAL I” 2 2 5*
Calibration function “CAL T” 2 2 6
* = factory setting
2–15