Printout/Record for Calibration and Linearization Functions (menu code 8 10 2)
A printout or record is generated only at the end of the following functions:
– all calibration operations
– all linearization operations
– air density determination (see part 3, “Application Programs”)
The printout of the record can have the following lines:
--------------------
MC1 - Sartorius : Balance family and manufacturer
Model MC5 : Balance model
S/N 030800046 : Serial no. of the balance
Id 4-32-1 : ID no. (e.g., for identifying
a workstation/operator)
--------------------
Date : 30-Jul-95 : Current date
Start: 10:05:30 : Time at which the application started
Cal. : Test :
Calibration mode (in this case, “calibration”)
Diff.:- 0.000013 g : Data generated during the calibration test
Cal. : Intern : Calibration mode (in this case,
“internal calibration”)
Stat.: Complete : Status comment for calibration or
linearization functions
End : 10:05:45 : End of application
Name : : Field for signature of the operator responsible
--------------------
Set. : 5.000000 g : Calibration weight (only for “external
calibration”)
Set. : 1.175 kg/ : Air density
Cal. : Air
2–26