Step/Key ID/Key Readout Data Output
Enter memory location/contianer
ID, e.g., 10 and store using F 10
Zero the balance using t
0.000 mg
Place container on pan; +
355.916 mg
store weight using F
0.000 mg
Net
Add sample to container; +
780.352 mg
Net
store weight using STO – F +
1136.268 mg
11
Remove container with sample
0.000 mg
Load next container; +
375.235 mg
store weight using - F
0.000 mg
Net
Add sample to container +
889.715 mg
Net
store weight using STO - F +
1264.950 mg
Store weights of other 12
containers and weigh-in
further samples
Treat samples (e.g., dry samples)
Enter 1 st memory location/
container ID 10 and 10
press to confirm - F
Zero the balance using t
0.000 mg
Place corresponding container +
934.696 mg
on pan; backweigh using +
74.17 %g
Net
Num 10
RE S - F Date 9-Jul-95
Time 10:34:54
Net + 780.352 mg
Res + 578.780 mg
Res + 74.17 %
If necessary, press w +
578.780 mg
Net
to display 2nd calculated value
Repeat procedure for all
samples RE S
v
- F
11
3–26