Practical Example: Density Determination Using a Specific Gravity Bottle
Reference liquid: Water
Temperature: 20°C
Balance used: MC 210 P
Menu code settings used in the example:
Function Code
Density 2 1 7
Tare memory 2 2 2
Selective c: f/F 2 4 3
Method: Spec. gravity bottle 2 6 2
Automatic output
of all calculated values 7 1 2
With data ID code 7 2 2
Auxiliary equipment: Specific gravity bottle with a capacity of 50 ml
Sartorius YDP 03-0CE Data Printer
Application: Determination of the density of powdery
or fine-grained samples
Procedure: Distilled water at 20°C will be used as the reference
liquid. The hard copy is to be identified by the sample
ID number “K* xxx.
After pouring out the reference liquid, you do not need to dry
the specific gravity bottle, if you store the weight of the
liquid remaining in the bottle temporarily in the tare memory.
Step/Key ID/Key Readout Data Output/Printout
To start program
clear functions: c f c F +
xxxxxx g
Enter sample ID no. 2.05:
2.05 i +
2.05 g
K* 2.05
Place empty spec. grav. bottle +
xx.xxxxx g
on pan; tare t
0.00000 g
Access program:
°C
STO - F +
20.0°C
As an alternative, you can select
this code:
Automatic output
of the density (sp. gravity) 7 1 3
3–40