Display Modes for Verified Balances Approved
for Use as Legal Measuring Instruments
You can select the display mode that best meets your individual requirements.
Last Numeral Blanked When the Load Changes
As the load on your balance changes, the display resolution is reduced by a
factor of 10 so that you will obtain a faster and more stable readout.
In the process, the last numeral is blanked until the load stabilizes. Once the
load has stabilized, the readout is shown again with the full display accuracy,
which means the last numeral is displayed.
Code
Display mode 1st range 2nd range 3rd range
Highest possible accuracy 1 8 1* 3 2 1*
Last numeral blanked
when load changes 1 8 2 3 2 2
Rounding factor 10 3 4 5*
IQ-Mode
(Load-Dependent Readability)
The IQ-mode does not apply to EC type-approved balances verified for use as
legal measuring instruments.
* = factory setting