Internal Linearization
Press the t key (26) for at least 2 seconds until
“C.I.” and “CAL” are displayed (next to the f key).
Select “internal linearization” by pressing the
F key (22) several times until “L.I.” is displayed.
Unload the balance and tare (close the draft shield).
“L.1.” stands for “internal linearization.”
When a zero readout is displayed, press the
f key (21). “C” will now be indicated. The built-in
weights are internally applied one after the other
by servomotor, and the balance is automatically
linearized.
If external interference affects the linearization
procedure, you may obtain a brief display of the error
message “Err 02.” in this case, tare and then press
the f key again when a zero readout appears.
An acoustic signal indicates the end of linearization.
Important Note:
The balance automatically self-calibrates after each
internal linearization procedure.
1–35