Perform a Rapid Run
90
Part # 15035786 Rev. D
Load a Flow Cell
The next step is to remove the used flow cell and load the flow cell that you want to
sequence.
NOTE
If the cBot was used for clustering, load the clustered flow cell. For on-instrument clustering,
load a new flow cell.
Remove the Used Flow Cell
1 Open the flow cell compartment door.
Figure 42 Flow Cell Lever in Position 1
2 Slowly move the flow cell lever to position 1 to disengage the manifolds.
3 Slowly move the flow cell lever to position 0 to disengage the vacuum seal and release
the flow cell.