Enter Run Parameters
HiSeq 2500 System User Guide
23
A first base report is generated automatically for each run. Selecting this option opens
the first base report before proceeding with the run.
2 [Optional] From the Align to PhiX checkboxes, clear the checkbox for lanes that do not
contain PhiX.
By default, all lanes are selected for alignment by Real-Time Analysis software.
Alternatively, select lanes on the flow cell image to add or remove lanes for PhiX
alignment.
NOTE
A dedicated control lane is not required with HCS v2.2 and RTA v1.18. Therefore, the option
to assign a control lane is not available with this software configuration.
3 Select Next.
Recipe Screen
1 Select from the following Index Type options:
• No Index—Performs a non-indexed single-read or paired-end run.
• Single Index—Performs a single-read or paired-end run with 1 indexing read.
• Dual Index—Performs a single-read or paired-end run with 2 indexing reads.
• Custom—Performs a single-read or paired-end run with a custom number of cycles
for index reads.
2 If the Dual Index or Custom option is specified, select a Flow Cell Format, either Single
Read or Paired End.
3 Enter the number of cycles for Read 1 and Read 2, if applicable.
NOTE
The number of cycles performed in a read is 1 more cycle than the number of cycles
analyzed. For example, to perform 125 cycles for Read 1, enter 126.
For the Custom indexing option, enter the number of cycles for index reads. Read
lengths do not need to be identical.
4 Confirm the following default chemistry settings. These fields are auto-populated
depending on the selected index type option.
a SBS: HiSeq SBS Kit v4
b Index: HiSeq v4 Single Index or HiSeq v4 Dual Index
c PE turnaround: HiSeq PE Cluster Kit v4
5 [Optional] Select the Use Existing Recipe checkbox to use a custom recipe. Otherwise,