Perform a TruSeq v3 Run
52
Part # 15035786 Rev. D
6 Place each reagent bottle onto the rack in the associated numbered position. Make sure
that the conical end of the bottle rests in the indentation on the base of the rack.
Position
Reagent Description
1
ICB Incorporation Mix
2
PW1 (25 ml) Wash Buffer
3
SRE Scan Mix Reagent
4
SBS Buffer 1 (SB1) High Salt Buffer
5
SBS Buffer 2 (SB2) Incorporation Wash Buffer
6
SBS Buffer 2 (SB2) Incorporation Wash Buffer
7
CMR Cleavage Mix Reagent
8
SBS Buffer 3 (SB3) Cleavage Buffer
Table 5 SBS Reagent Positions
7 Add 25 ml of PW1 or laboratory-grade water to the bottle in position 2.
8 Slide the reagent rack into the reagent compartment, aligning the rack with the raised
guide on the floor of the compartment.
9 Lower the sippers into the sequencing reagent bottles as follows:
a Pull the sipper handle towards you and then lower the sipper handle.
b Visually inspect the sippers to make sure that they do not bend as they lower into
the funnel caps.
c Release the sipper handle into the slot on the bottom end of the groove.
10 Select the PW1 (25 ml) loaded checkbox.
Load Indexing Reagents
1 Record the weight of each reagent on the lab tracking form.
2 Make sure that the paired-end rack is not in use on the adjacent flow cell. Steps that
use the paired-end rack include Read2 resynthesis, Index 1 (i7) Read preparation, and
Index 2 (i5) Read preparation.
3 Raise the sippers for the paired-end reagent rack using the following motion:
a Pull the handle towards you and then raise the handle.
b Release the handle into the slot on the top end of the groove. Make sure that the
handle rests securely in the slot.