Load and Prime Reagents
HiSeq 2500 System User Guide
81
Load and Prime Reagents
After entering run parameters, load SBS, clustering, indexing, and paired-end reagents for
the run, and then prime reagents through the fluidics system. The software guides you
through these steps in a series of screens on the Pre-Run Setup tab.
Illumina-Supplied Consumables
} Eight funnel caps
User-Supplied Consumables
} 250ml bottle (Corning, catalog#430776)
} 15 ml conical tubes (Corning, catalog # 430052)
} Laboratory-grade water
NOTE
To prepare for the post-run rinse at the end of a sequencing run, load 25 ml PW1 or
laboratory-grade water in position 2.
The post-run rinse does not replace the post-run instrument wash.
Load SBS and Clustering Reagents
Make sure that SBS reagents are ready to load onto the instrument.
Illumina-Supplied Consumables
} Eight funnel caps
User-Supplied Consumables
} One 250ml bottle (Corning, catalog#430776)
} One flip-top Eppendorf tube per flow cell (1.5 ml or 1.7 ml). Do not use tubes with
screw caps.
Procedure
1 Record the weight of each reagent on the lab tracking form.
2 Open the reagent compartment door.
3 Raise the sippers for the sequencing reagent rack using the following motion: