Perform a Maintenance Wash
HiSeq 2500 System User Guide
103
17 [For Rapid Run mode] Place the ends of tubes 4 and 5 into an empty container. Place
the end of all of the other tubes into a bottle of clean water to prevent air from being
introduced into the syringe pumps.
18 Select Next to start the wash.
19 When the wash is complete, select Return to Start.
20 Measure the delivered volume.
NOTE
All bottles and tubes are filled to capacity to make sure that the sippers are rinsed. However,
the delivered volume for each position varies so the bottles and tubes contain different
volumes when the wash is complete.
Positions High Output Flow Cell
Delivered Volume
Rapid Flow Cell Delivered
Volume
8 SBS positions 82 ml 29ml
10 PE positions 76 ml 30ml
1 library position Empty 1.2 ml
All positions 19.75 ml per lane 30.1 ml per lane
21 Unwrap the waste tubes and return them to the waste container.