Directed Electronics equalizer 2500 Guía de usuario Pagina 32

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 154
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 31
Perform a HiSeq v4 Run
22
Part # 15035786 Rev. D
analysis, this field is optional.
2 Select Zip BCL files to reduce required storage space. If the run is connected to
BaseSpace, the Zip BCL files option is selected by default.
NOTE
The Bin Q-Scores setting is enabled by default to reduce required storage space. This setting
groups quality scores over a wider range of values without affecting accuracy or
performance.
3 Select from the following Save Auxiliary Files options:
Save All Thumbnails—Saves all thumbnails images. A thumbnail is a sampling of
images from many tiles in each column of tiles, or swath, combined in 1 thumbnail
image.
Save Tile Thumbnails—Saves tile thumbnails. Tile thumbnails represent a single
tile rather than a sampling of tiles in a swath.
4 Select Next.
Flow Cell Setup Screen
The Flow Cell Setup screen records information about the flow cell used for the run. All
fields are required.
1 Scan the flow cell barcode or enter the flow cell ID (barcode number) of the flow cell to
be sequenced. The flow cell ID is used to determine flow cell type and reagent
compatibility.
2 Confirm that the flow cell type is HiSeq Flow Cell v4. The flow cell type is selected
automatically based on the flow cell ID.
3 Enter an experiment name. The experiment name appears on each screen to help
identify the run in progress.
4 Enter a user name.
5 Select Next.
Advanced Screen
1 [Optional] Select the Confirm First Base checkbox.
Vista de pagina 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 ... 153 154

Comentarios a estos manuales

Sin comentarios