Load a Flow Cell
HiSeq 2500 System User Guide
91
Figure 43 Flow Cell Lever in Position 0
4 Lift the used flow cell from the flow cell holder.
Clean the Flow Cell Holder
1 Put on a new pair of powder-free latex gloves.
2 Using an alcohol wipe or a lint-free tissue moistened with ethanol or isopropanol,
carefully wipe the surface of the flow cell holder until it is clean.
CAUTION
Do not allow alcohol to drip into the vacuum holes or around the manifolds. Use a low-lint
lab tissue to dry the stage, if necessary.
Figure 44 Inspect Vacuum Holes
3 Visually inspect the flow cell holder to make sure that it is free of lint and the vacuum
holes are free of obstructions.