Post-Run Procedures
108
Part # 15035786 Rev. D
Idle the Instrument
Use the following instructions to prepare the instrument to sit idle for up to 10 days. For
durations longer than 10 days, see Shut Down the Instrument on page 109.
1 Perform a complete maintenance wash to flush the system fully. For more information,
see Perform a Maintenance Wash on page 100.
2 Leave the flow cell on the flow cell stage with the flow cell lever in position2. The
manifolds remain in the raised position.
3 [For high output modes] Load 10ml laboratory-grade water in each reagent position in
the reagent racks. Then lower the sippers.
4 [For Rapid Run mode] Load 10ml laboratory-grade water in each reagent position in
the reagent racks and 1 ml laboratory-grade water in the loading station position. Then
lower the sippers.
5 Do not turn off the instrument.
6 Before using the instrument again, perform a water wash. For more information, see
Perform a Water Wash on page 104.