Troubleshooting
136
Part # 15035786 Rev. D
Pause a Run
CAUTION
Do not pause a run during imaging. Use the Normal stop, end of cycle or end of lane feature,
to stop and resume a run.
Pause a run from the Run Overview screen. Pausing a run might be necessary to check run
components, such as reagent volumes, before proceeding with the run. Under normal
operation, pausing a run is not necessary.
1 From the Run Overview screen, select Pause. The pause menu opens.
Figure 55 Pause Options
2 Select Normal Pause.
3 Select Yes to confirm the pause command. The software completes the current
chemistry or imaging command and places the flow cell in a safe state.
4 Select Resume to resume the run.
Change Reagents During a Run
If you started the run with a partial volume of reagents, use the Change Reagents feature to
pause the run and replenish reagents.
1 From the Run Overview screen, select Pause. The pause menu opens.
2 Select Change Reagents.
3 Select Yes to confirm the pause command. The software completes the current
chemistry or imaging command, places the flow cell in a safe state, and opens the
Reagents screen.
4 On the Reagents screen, enter the following reagent parameters:
The reagent kit ID for new reagents.
The number of cycles the reagents are expected to last.
NOTE
The priming checkbox is disabled on the Change Reagents screen. Priming is not required.
5 Select Next to proceed to loading reagents.